CONTROLLER SPECIFICATIONS
BASE TEMPERATURE RANGE Continuously adjustable over range 20 - 30 C with flat
bladed screwdriver
CONTROL RANGE 20.5either side of nominal base temperature point in
switchable 0.1 steps. 45 C max.
CONTROL ACCURACY 0.1 C
DIGITAL READOUT: Resolution 0.1 C
Accuracy 0.1, digit
AMBIENT OPERATING RANGE 15 - 45 C
INPUT POWER REQUIREMENTS 100 - 120V AC, 60Hz, 100W
200 - 240V AC, 50Hz, 100W
SIZE 8" high x 7" wide x 15" deep
WEIGHT 28 lbs
OTHER FEATURES Spring loaded switch indication of set-point Safety
shutdown with warning lamp in case of fault condition
such as thermocouple sensor breakage, lack of cooling
water or electronics failure. Self-sealing water connectors.
Auxiliary AC switched output.
THERMAL PROBE SPECIFICATIONS
SIZE 6.5” long x 5/8” dia. Total length with cable, 6 feet approx
WEIGHT 0.5 lb
TEMPERATURE CONTROLLED 0.5” diameter
SURFACE AREA
RANGE 0 - 45 C
Note: Sustained exposure to temperatures over 46 C can
cause burns. This equipment is not intended for use at
temperatures over 45 C
RESPONSE TIME Less than 4 seconds in heating or cooling
OTHER FEATURES Built-in feedback sensor, type T, copper-constantan.
Self sealing water connectors.